Optris Logo - Affordable temperature measurement solutions
tune

Product Configurator

Finding the perfect Optris product for your needs.

IR Thermometer Configurator arrow_forward IR Camera Configurator arrow_forward
Configurator Screen
lens_blur

Optics Calculator

Use the Calculator to quickly determine the right spot size for your needs.

Explore Now arrow_forward
Calculator Screen
forum

Chat with Engineers

Online service support

Opening Hours
Monday – Friday:
8:00 am – 5:00 pm

 

Chat Now arrow_forward
call

Call Us

Sales department:
+1 603 766-6060

Opening Hours
Monday – Friday:
8:00 am – 5:00 pm

 

mail

Email Us

Sales department:
[email protected]

Contact Us arrow_forward
construction

Request a Repair

Service request for repair orders:
[email protected]

Service Request Form arrow_forward

Calibration Source BR 500

Black Body for Infrared Sensor Calibration

Calibration Source BR 500
Calibration Source BR 500
Calibration Source BR 500
Calibration Source BR 500

The Optris BR500 is a high-performance infrared reference radiation source (blackbody) for precise calibration of infrared cameras and pyrometers. As the successor to the BR400, it offers improved usability, enhanced performance, and extended temperature capability.

With a maximum reference temperature of up to 500 °C, the BR500 expands the application range for verification and calibration tasks. It delivers highly stable and homogeneous radiation for reliable, repeatable results. Accurate radiation temperature measurement can be performed using either a contact thermometer with emissivity determination or a transfer standard infrared thermometer.

A modern touch display enables fast setup and intuitive operation. Its compact, rugged design, combined with high precision and long-term stability, makes the BR500 ideal for both initial calibration and routine verification.

  • Radiator temperature up to 500 °C
  • Excellent homogeneity, precision & long-term stability
  • Compact and rugged design
  • Perfect for calibrating and testing infrared sensors

$7,500.00

excl. sales taxes, custom tariffs and shipping

Product Description

The Optris BR500 is a high-performance infrared reference radiation source (blackbody) for precise calibration of infrared cameras and pyrometers. As the successor to the BR400, it offers improved usability, enhanced performance, and extended temperature capability.

With a maximum reference temperature of up to 500 °C, the BR500 expands the application range for verification and calibration tasks. It delivers highly stable and homogeneous radiation for reliable, repeatable results. Accurate radiation temperature measurement can be performed using either a contact thermometer with emissivity determination or a transfer standard infrared thermometer.

A modern touch display enables fast setup and intuitive operation. Its compact, rugged design, combined with high precision and long-term stability, makes the BR500 ideal for both initial calibration and routine verification.

Specification

General Specifications
Operating temperature: 5 °C … 30 °C
Weight: 9.8 kg
Dimensions (H x W x D): 380 x 240 x 230 mm

Calibration Specifications
Temperature range (at 23 °C): +5 °C … 500 °C
Temperature resolution: 0.1 °C
Emissivity: 0.97 ±0.02 (for 8 –14 μm)
Aperture: 128 mm

Accuracy:
± 0.35 °C at 35 °C
± 0.50 °C at 100 °C
± 0.70 °C at 200 °C
± 1.50 °C at 350 °C
± 1.9 °C at 500 °C

Warm-up time:
20 Min. (from 35 °C to 450 °C)
10 Min. (from 450 °C to 500 °C)

Cool-down time:
40 Min. (from 500 °C to 100 °C)
40 Min. (from 100 °C to 35 °C)

Electrical Specifications
Temperature sensor: Pt100
Controller: PID
Power supply: 220V/ AC 50 Hz (±10 %) (optional: 110V/ AC 60 Hz model)
Power consumption: Max. 1000 W

Downloads
FAQs

What should be the distance between the pyrometer and the calibration source?

Each infrared sensor is calibrated at a specific distance, depending on the device and its optics. Please, contact Optris for the correct calibration geometry.

How do I calibrate an infrared sensor?

The calibration process typically involves adjusting the sensor’s settings to match a known temperature, usually provided by a calibration source. This ensures the sensor’s readings are accurate. For specific instructions, please consult the device’s manual or contact Optris. A reference pyrometer is most likely required.

 

Scope of Supply
  • Calibration source
  • Power supply cable
  • Calibration certificate